Skip to main content
← Medical Physics

Medical Physics · 29.1

X-Ray Imaging

Slam a beam of fast electrons into a tungsten block and about one percent of their energy re-emerges as X-ray photons — the other ninety-nine percent is heat. Aim those photons through a patient at a detector, and the shadow the skeleton casts becomes the oldest medical image of all.

01

Build the model

Connect the measurement to the mechanism.

An X-ray tube boils electrons off a hot cathode by thermionic emission, accelerates them through a p.d. of tens of kilovolts in vacuum, and crashes them into a tungsten anode. The violent deceleration radiates a continuous braking spectrum whose sharp short-wavelength cutoff comes from an electron converting its entire kinetic energy into one photon: eV = hc/λmin. Electrons that knock out inner tungsten electrons add characteristic spike lines at fixed wavelengths on top of the continuum.

Only about 1% of the beam power becomes X-rays, so the anode spins to spread the heat. Inside the body the beam attenuates exponentially, I = I₀e^(−μx), and the attenuation coefficient μ rises steeply with atomic number and falls with photon energy: calcium-rich bone absorbs far more than soft tissue, so the transmitted beam prints a shadowgraph. Soft structures that barely differ in μ are made visible by filling them with a high-Z contrast medium such as a barium meal.

Simple definition
An X-ray tube converts the kinetic energy of electrons accelerated through a p.d. V into X-ray photons — maximum photon energy eV, so λmin = hc/eV. The body attenuates the beam as I = I₀e^(−μx), and differences in μ between tissues create the image contrast.
Example
At 100 kV no photon can carry more than 100 keV, so λmin = hc/eV ≈ 12.4 pm. Raising the voltage shifts the whole spectrum to shorter, more penetrating wavelengths; raising the current only makes more of the same photons.
Minimum wavelengtheV = hc/λmin

The most extreme electron gives everything to a single photon. Nothing in the tube can beat that, so the spectrum stops dead at λmin — a cutoff you can move with the voltage knob.

λmin = hc/eV — set by the tube p.d. alone

Exponential attenuationI = I₀e^(−μx)

Each layer of tissue removes the same fraction of whatever intensity reaches it — so intensity falls exponentially, never to zero, exactly like radioactive decay with depth playing the role of time.

μ is the linear attenuation coefficient

Electron energyEₖ = eV

A 100 kV tube gives each electron 100 keV. Tungsten is chosen for its high Z (efficient X-ray production) and high melting point (it must survive the other 99%).

≈ 99% ends as heat in the tungsten target

Contrastμ rises steeply with Z

Bone is rich in calcium (Z = 20) while soft tissue averages Z ≈ 7 — that Z³ gap is the whole radiograph. Barium (Z = 56) borrows the same trick to paint the gut visible.

roughly as Z³ at diagnostic energies

01

Inside the tube

A heated filament releases electrons (thermionic emission); a p.d. of 30–150 kV accelerates them across an evacuated gap; a tungsten anode stops them. The tube current sets how many electrons arrive per second and hence the beam intensity; the tube voltage sets each photon's energy budget. Because ~99% of the power becomes heat, the anode is a spinning disc, presenting fresh metal to the beam thousands of times a second.

02

Two spectra in one beam

The continuous spectrum comes from braking radiation: electrons decelerating in the target radiate every wavelength down to the λmin cutoff. The line spectrum comes from tungsten itself: a beam electron ejects an inner-shell electron, an outer electron falls into the hole, and a photon of fixed, element-specific energy leaves. The lines only appear once the tube voltage exceeds the inner-shell binding energy — about 70 kV for tungsten's K lines.

03

From shadow to image

Sharpness demands a small focal spot and a patient close to the detector, so each point casts a crisp shadow instead of a blurred penumbra. Contrast demands tissues of different μ in the path — given for bone, engineered with barium for the gut. An aluminium filter hardens the beam by absorbing the long-wavelength photons that would only dose the skin without ever reaching the detector, and digital flat-panel detectors have replaced film, trading grains of silver for pixels and slashing the dose needed.

02

Change one variable at a time

Make the relationship visible.

Raise the p.d. and the cutoff λmin = hc/eV slides left — plus, past ≈ 70 kV, tungsten’s characteristic spikes switch on. The filter eats the long-wavelength tail that would only dose the skin. Current scales the whole curve but never moves the cutoff.

λmin = 15.5 pmK linesI(λ)λ (pm)

Cutoff λmin = hc/eV15.5 pm · max photon 80 keV

Characteristic linesexcited — p.d. beats the K-shell binding

Beam power P = VI16 kW

X-rays ≈ 1% · heat ≈ 99%160 W out · 15.8 kW of heat

03

Catch the common trap

Explain before calculating.

Doubling an X-ray tube's accelerating p.d. does what to the minimum wavelength λmin?

Choose an answer to test the model.

04

Worked examples

State the rule, substitute, then check units.

EasyFind the minimum X-ray wavelength produced by a tube operating at 80 kV.
  1. The most energetic photon carries the whole electron energy: eV = hc/λmin.
  2. λmin = hc/eV = (6.63 × 10⁻³⁴ × 3.00 × 10⁸) ÷ (1.60 × 10⁻¹⁹ × 8.0 × 10⁴).
  3. λmin = 1.55 × 10⁻¹¹ m ≈ 16 pm — deep in the X-ray band.

Answerλmin ≈ 1.6 × 10⁻¹¹ m (16 pm)

MediumA beam passes through 5.0 cm of muscle with μ = 0.20 cm⁻¹. What fraction of the intensity emerges?
  1. I/I₀ = e^(−μx) = e^(−0.20 × 5.0) = e^(−1.0).
  2. e^(−1.0) = 0.37.
  3. About 37% of the intensity survives — attenuation removes a fixed fraction per centimetre, so the answer would be 0.37² ≈ 14% for 10 cm.

Answer≈ 37% transmitted

HardOne ray crosses 4.0 cm of soft tissue (μ = 0.22 cm⁻¹). A neighbouring ray crosses 3.0 cm of tissue plus 1.0 cm of bone (μ = 0.60 cm⁻¹). Find the intensity ratio between the two rays — the contrast in the image.
  1. Tissue-only ray: I₁/I₀ = e^(−0.22 × 4.0) = e^(−0.88) = 0.415.
  2. Tissue + bone ray: exponents add along the path: I₂/I₀ = e^(−0.22 × 3.0 − 0.60 × 1.0) = e^(−1.26) = 0.284.
  3. Ratio I₂/I₁ = 0.284 ÷ 0.415 = 0.68 — the bone path arrives about 32% dimmer, and that difference is the shadow the radiologist reads.

AnswerI₂/I₁ ≈ 0.68 — one centimetre of bone dims the ray by about a third

ChallengingA tube runs at 90 kV and 250 mA. Find the electrical power, the approximate X-ray output, the number of electrons striking the target per second — and explain the rotating anode.
  1. P = VI = 9.0 × 10⁴ × 0.250 = 22.5 kW of electron beam power.
  2. At ~1% efficiency, only ≈ 225 W leaves as X-rays; ≈ 22.3 kW becomes heat in a focal spot barely a millimetre across.
  3. Electrons per second = I/e = 0.250 ÷ 1.60 × 10⁻¹⁹ = 1.6 × 10¹⁸ s⁻¹.
  4. No stationary target survives 22 kW on a millimetre: the anode is a disc spinning at thousands of rpm, so each part of the rim is heated only briefly and radiates its heat between passes.

Answer22.5 kW in, ≈ 225 W of X-rays out, 1.6 × 10¹⁸ electrons per second — hence the spinning anode